Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: ©Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2023, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 22 July 2022
Date accepted: 21 November 2022
Publication date (print and electronic): 15 February 2023
Volume: 33
Issue: 1
Electronic Location Identifier: 011001
Publisher ID: bm-33-1-011001
DOI: 10.11613/BM.2023.011001
Fibrin clot interference in a human chorionic gonadotrophin assay causing a false Down syndrome screening result
Arzu Etem Akagac[*]
Author notes:
[*] Corresponding author: arzuetem@yahoo.com
Serum samples are generally used for the measurement of human chorionic gonadotrophin (hCG) to calculate second-trimester maternal screening results. Lower hCG concentrations correlate with a lower calculated risk of Down syndrome (DS). Hence, erroneously low hCG results due to fibrin clot may lead to misinterpretation. We present a 23-year-old woman with a pregnancy of 17+3 weeks. Blood was taken into the Becton-Dickinson (BD) vacutainer SST-II Advance tube (Ref: 367955). The hCG test was performed on Immulite 2000 XPi analyser (Siemens Healthcare Diagnostics Inc, Tarrytown, USA) with original reagents. The results of the same sample were found as 2566 U/L, 18,153 U/L, and 7748 U/L. Three consecutive results after removal of the small fibrin clot and recentrifugation were 18,878, 20,255, and 22,339 U/L. The risk of DS and MoM for the concentration of 2556 U/L hCG was < 1/10,000 and 0.14, respectively. For a hCG concentration of 20,255 U/L, these values were 1/5632 and 1.13, respectively. Laboratory professionals and technicians should be aware that erroneously low hCG results can be measured with the Immulite 2000 XPi due to interference from small fibrin clots. Falsely underestimated hCG values reduce the MoM values and thus the calculated risk of DS.
Keywords: hCG; prenatal screening; fibrin; preanalytical phase; case report