Journal Information
Journal ID (publisher-id): BM
Journal ID (nlm-ta): Biochem Med (Zagreb)
Title: Biochemia Medica
Abbreviated Title: Biochem. Med. (Zagreb)
ISSN (print): 1330-0962
ISSN (electronic): 1846-7482
Publisher: Croatian Society of Medical Biochemistry and Laboratory Medicine
Article Information
Copyright statement: ©Croatian Society of Medical Biochemistry and Laboratory Medicine.
Copyright: 2020, Croatian Society of Medical Biochemistry
License (open-access):
This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Date received: 13 January 2020
Date accepted: 21 March 2020
Publication date (print and electronic): 15 June 2020
Volume: 30
Issue: 2
Electronic Location Identifier: 020707
Publisher ID: bm-30-2-020707
DOI: 10.11613/BM.2020.020707
The results of external quality assessment programme on urine leukocyte and erythrocyte counting in Poland
Introduction
Urine particle analysis is an important diagnostic tool. The aim of this study was to evaluate the quality of urine leukocyte (WBC) and erythrocyte (RBC) counting results obtained with manual and automated methods in Polish laboratories participating in the international external quality assessment (EQA) programme.
Materials and methods
1400 WBC and RBC counting results were obtained from 183 laboratories in EQA surveys organised by Labquality (Helsinki, Finland) from 2017 to 2019. The between-laboratory coefficient of variation (CV), the percentage difference between the laboratories' results and target values (Q-score (%)), as well as modified Youden plots were analysed.
Results
For automated method groups, the medians of inter-laboratory CVs varied from 14% to 33% for WBC counting and from 10% to 39% for RBC counting. For manual method groups, the medians of CV varied from 53% to 71% (WBC) and from 55% to 70% (RBC), and they were significantly higher, in comparison to CVs for most automated method groups (P < 0.001). The highest percentage of results outside the target limits (36%) and the highest range of Q-score (%) (from - 93% to 706%) were observed for laboratories which participated in the surveys for the first or second time. The percentage of deviating results and the ranges of Q-score decreased with an increased frequency of laboratories’ participation in the surveys.
Keywords: erythrocytes; leukocytes; quality control; standardisation; urine